EasyManua.ls Logo

AVL 9180 - Page 75

AVL 9180
114 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
69
ERROR: UPPER NEEDLE SENSOR
If probe movement to the upper position is not detected within a certain time
period, the above message is displayed.
Possible remedies:
Check for foreign material that may prevent the probe from moving freely.
Contact AVL for technical assistance.
ERROR: LOWER NEEDLE SENSOR
If probe movement to the lower position is not detected within a certain time period,
the above message is displayed.
Possible remedies:
Make sure the fill port is installed correctly.
Check for foreign material that may prevent the probe from moving freely.
Contact AVL for technical assistance.
WHEEL MISSING OR SAMPLER DEFECT
During the homing sequence, the sampler checks for the sample wheel to be
present.
Possible remedies:
Make sure the wheel is present and seated correctly.
Replace sample wheel.
Replace sampler.

Table of Contents

Other manuals for AVL 9180

Related product manuals