14 USER MANUAL
____________________________________________________________________________________________________________________
In order to perform internal calibration proceed with the following:
Empty the pan.
Press key twice (double pressing the key
helps to avoid accidental starting calibration
procedure).
During calibration internal weight is put three
times on and obtained results are compared.
Discrepancy of results is signalled with a
message and causes the balance being blocked.
Until calibration process is finished do not
perform any operation on the balance. Any
vibrations and shocks interfere calibration
process and may delay it or deteriorate accuracy
of its result.
When internal calibration is performed
successfully the balance indicates zero on the
display at empty pan.
Note:
In order to terminate calibration process in not
legally verified balances press key and wait
until balance mechanism is not settled in initial
position.
g
g
~45 seconds