ULT FREEZER – USER MANUEL
10
Updated 19. April 2021
Temperature control ULT Freezers.
The temperature inside the Freezer is controlled by the electronic controller in the top panel.
The controller has a digital readout of the temperature inside the cabinet, and the option of
changing the temperature inside the cabinet.
Backup Battery. (For temperature logging and Alarm, during Power cut)
The battery should be exchanged every two years.
Please note! When commissioning the Freezer, the Battery must be checked for correct
connection.
Controller
1. Main Menu
DISPLAY SHOWS
1. Date
2. Time
3. Temperature inside the cabinet
4. Room Temperature (Ambient)
5. Shows if compressor is running
(it will not be visible if the
compressor is not running)
6. Compressor is running/defrost
7. USB is connected
8. Battery status
DISPLAY GIVES ACCESS
TO
9. Temperature graph
10. Status
11. Download
12. Menu
13. ON/OFF
1
2
3
4
6
7
5
8
9
10
13

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the B SCIENCE B-ULT 100 and is the answer not in the manual?

B SCIENCE B-ULT 100 Specifications

General IconGeneral
BrandB SCIENCE
ModelB-ULT 100
CategoryFreezer
LanguageEnglish