ULT FREEZER – USER MANUEL
37
Updated 19. April 2021
1. Alarm List
Under alarm list it is shown which
type of alarms are occurring. The
Button will be red when the failure
is activated, and the remote alarm
relay will be active until you click
on the red button to deactivate the
alarm.
Pay special attention if you get a Compressor alarm
:
The controller has switched OFF the Compressor relay, and the unit must be restarted, by
Switching OFF and then ON again to make a reboot. If this Compressor alarm occurs
again, contact a technician to check why the Compressor alarm occurs.
Under the Sensor menu.
Sensor Menu
When a Sensor alarm occurs, you can see which alarms are in failure.
Click on the red button to deactivate the alarm.