Introduction ECA 450
2 Instruction 0024-9400
Performs data logging. A series of combustion tests can be automatically
started at user-defined intervals, and the data saved for reviewing at the
conclusion of the logging period. The data can be stored in either the
analyzer’s internal memory, sent to a personal computer, or both.
Utilizes four sources of primary-air temperature information in the
following priority: 1) A thermocouple can be inserted into the primary-air
stream and plugged into the analyzer’s front panel T-AIR connector. 2) A
primary-air temperature reading can be made with the probe’s
thermocouple and stored in memory. 3) An external thermometer can be
used to take a primary-air temperature reading and then entered into
memory using the analyzer’s front panel keypad. 4) If no other primary-air
temperature source is available, the analyzer defaults to an internal case
temperature that is measured the on printed circuit board.
Stores test identification that is manually entered by an operator. This
information is added to the test records as they are saved, and appears
within each test record when printed.
Stores user name information that is manually entered by the owner or
user of the analyzer. This information appears at the top of each printout.
Displays information in either English, French, or Spanish.
Can be set up to automatically turn off the front-panel display to conserve
battery life.
1.3 Technical Data
The ECA 450 Directly Measures and Displays:
Oxygen content in flue gas in the range of 0.1 to 20.9% O2.
Stack gas temperature in the range of 4 to 2400 ºF (20 to 1315 ºC).
Primary-air in the range of 4 to 999 ºF (20 to 999 ºC).
Pressure in the range of ±27.7 inches of water column (±69 mb).
Carbon Monoxide content in flue gas in the range of 0 to 4,000 ppm CO
(corrected for the presence of Hydrogen).
Optionally . . .
Carbon Monoxide in the range of 4,000 to 80,000 ppm CO.
Nitric Oxide in the range of 0 to 3,500 ppm NO.
Nitrogen Dioxide in the range of 0 to 500 ppm NO
2.
Sulfur Dioxide in the range of 0 to 4,000 ppm SO
2.
Combustibles in the range of 0 to 5% gas.
The ECA 450 Calculates and Displays:
The following is true only when the measured Oxygen level is not above 16.0%,
and the Stack (Flue Gas) temperature is not above 2192 °F (1200 ºC)
Combustion efficiency in the range of 0.1 to 100%.
Excess Air in the range of 1 to 250%.
Carbon Dioxide content in flue gas in the range of 0.1 to a fuel dependent
maximum CO
2
value in percent.
Carbon Monoxide content referenced to a user defined percentage of
Oxygen in the range of 0 to 99,999 ppm CO.
Differential pressure in the range of ±27.7 inches of water column
(±69 mb).