ECA 450 Maintenance
Instruction 0024-9400
67
4.4.3 HC Sensor Replacement
Unlike the other sensors that are held in place by their associated bayonet
holders, the HC sensor is plugged into its printed circuit board which, in
turn, is secured inside the sensor compartment with four screws. Remove this
board and replace the sensor as follows:
Material Required:
Medium Phillips screwdriver
Small flat blade screwdriver
Replacement HC sensor (refer to Section 6.1)
Procedure:
1. Gain access to sensor compartment by first removing two screws from the
sensor-cover plate, and then removing plate (see Figure 4-3).
2. Unplug the HC sensor’s printed circuit board from its connector jack.
3. Remove the sensor’s printed circuit board by first removing the board’s
four corner screws, and then lifting the board with the HC sensor
attached out from the sensor compartment.
4. Unplug old sensor and discard.
5. Remove replacement sensor from its plastic canister; then plug sensor
into its printed circuit board.
6. Install and secure the HC sensor’s circuit board inside the sensor
compartment using the screws removed in Step 3.
7. Install sensor-cover plate that was removed in Step 1.
8. Place analyzer in an area of fresh air and turn it ON. After the warm-up
cycle completes, observe that no sensor error messages should have
appeared.
9. Calibrate the new sensor per Section 3.