Operating Manual BaehrTec S2000
Page 46/52 EnglishV101 Status 07/2015
Fault
Cause
Solution
Instruments
do not
hold
Maximum permitted speed
for instrument in use is
exceeded
Instrument shaft worn
Instrument shaft bent
Instrument not correctly
inserted in clamping
mechanism
Cream, ointment, or similar
on instrument shaft
To much work pressure on
instrument
Pulling movement while
working
Only work with the
maximum permissible
speed for the instruments
used
Check the instrument shaft
and use a new instrument
if necessary
Replace instrument
without fail
Insert instrument until stop
in clamping mechanism
Clean instrument. Work
with small instruments
until larger instruments
start to clamp securely
again. During operation,
always make sure that the
handpiece tip and the
instruments are clean.
Reduce pressure
Avoid pulling movements
Speed only
possible up to
10,000 rpm.
Device
automatically
switches back
to 10,000 rpm.
Insufficient spray solution in tank.
The device switches back to
10,000 rpm to prevent the
handpiece from getting warm.
Refill spray liquid
Handpiece
vibrates;
Handpiece is
loud with used
instruments
Maximum permitted speed
for instrument in use is
exceeded,
Instrument defective (bent,
shaft worn, etc.)
Only work with the
maximum permissible
speed for the instruments
used
Use new instruments.
Push button on
handpiece
blocked or
is difficult to
activate
Handpiece or handpiece button
soiled
Clean button
If one of these faults still persists despite having followed this information, please send the
unit to the customer service team.
In addition, please get in touch with the manufacturer for each fault / each malfunction in
cases of doubt.