EasyManua.ls Logo

Baker BDINQ400 - Preparing the SA Module and Cabinet for Installation; Step 1: Removing Side Covers and Panels

Default Icon
15 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
398D003 Rev C 2 of 15
II PREPARING THE SA MODULE & CABINET FOR INSTALLATION
Step 1: Remove the right side duct cover and lower SA module panels to gain access to the mounting holes
for the BDINQ400 cabinet on the SA chassis underneath the table top.
Remove front grille.
Step 2: Detach slide warmer from table top by removing mounting screws located underneath the tabletop
inside the bottom plenum. Move the slide warmer forward in order to clear the area in front of the printer
module. Slide the printer to the rear of the instrument.
Note: if this a field upgrade installation, remove existing touch screen, printer cover, front shifter panel and all
associated wiring for the touch screen. Install new front shifter panel p/n 200000025 that is designed for
cabinet interface.
Slide warmer
Front shifter panel
Duct cover
Front grille