Interface IO-Link
Données de paramétrage
Index
hex
Subindex
hex
Object name Access Length Value Range Gra-
dient
Unit Default
0x3C 0x01 SetPoint 1 = SP1 R/W 2 octets Process Data 100%
0x3C 0x02 SetPoint 2 = rP1 R/W 2 octets Process Data 0%
0x3E 0x01 SetPoint 1 = SP2 R/W 2 octets Process Data 100%
0x3E 0x02 SetPoint 2 = rP2 R/W 2 octets Process Data 0%
0x60 0x00 Password W 4 octets 0000..9999 0
0xD0 0x00 Delay Switching Time 1 R/W 2 octets 0..500 0.1 sec 0
0xD1 0x00 Delay Back Switching Time 1 R/W 2 octets 0..500 0.1 sec 0
0xD2 0x00 Delay Switching Time 2 R/W 2 octets 0..500 0.1 sec 0
0xD3 0x00 Delay Back Switching Time 2 R/W 2 octets 0..500 0.1 sec 0
0xD5 0x00 Min Pressure Value R 2 octets Process Data
0xD6 0x00 Max Pressure Value R 2 octets Process Data
0xD7 0x00 Measure damping R/W 2 octets 0…1000
in 10 ms steps
1 ms 0
Capteurs de pression avec IO-Link
BSP-B... /-V...