Technical data
Barkey GmbH & Co. KG - Gewerbestrasse 8 - 33818 Leopoldshoehe - Germany - Telephone: +49(0)5202-9801-0 31
Service department: +49(0)5202-9801-30 Fax: +49 (0)5202-9801-99 - email: info@barkey.de
Instructions for Use
Barkey S-line
9 Technical data
Manufacturer Barkey GmbH & Co. KG
Gewerbestrasse 8
33818 Leopoldshoehe
Germany
Tel.: +49(0)5202-9801-0
Fax: +49(0)5202-9801-99
Type Barkey S-line
Article number 213.10285
Connection 100–240 VAC, 50–60 Hz
Wattage max. 75 watts
Power consumption typ. 1.0 A at 230 VAC
typ. 1.8 A at 115 VAC
Fuse 2.5 A time-delay
Temperature setting +39°C
Control precision ± 2.0°C
Overtemperature protection
warming section 43.0°C ± 1.0°C
Warming up time From 20°C to 39°C approx. 2 min.
Operations Storage/transport
Ambient temperature +10°C to +40°C -20°C to +70°C
Relative humidity 30% to 75% 30% to 90%
(non-condensating)
Air pressure 700 to 1060 hPa 700 to 1060 hPa
Mode Permanent operations
Dimensions Control unit W x D x H: 90 x 60 x 160mm
Warming section for infusion lines Ø5mm, length 1.5m
Weight approx. 1.3 kg
Safety class I protective earthing
Degree of protection BF; defibrillation-protected
External and internal use on patients
with insulated application part
MDD classification II a
Moisture protection IP X2
Marking CE 0123