Hygienic maintenance disinfection schedule
The hygienic maintenance of the machine must follow this schedule;
4/29/20
109
Frequency
Activity
Result
After each
treatment
1. Run a heat disinfection program (with or without citric acid), or a short heat citric program.
2. Wipe the outside of the dialysis machine with 70% ethanol or 60% isopropanol.
3. Rinse the outside and flush the inside of the pick-up tubes with water. Let them dry naturally.
Disinfection
At least after every
3
rd
treatment
1. Run a heat disinfection program together with CleanCart C cartridge. Cleaning
Decalcification
Disinfection
At least once
every 7
th
treatment
day
1. Run a chemical disinfection program with sodium hypochlorite.
A heat disinfection program together with CleanCart C cartridge shall be performed before a
sodium hypochlorite program.
or
1. Run a heat disinfection program together with CleanCart C cartridge.
2. Run a heat disinfection program together with CleanCart A cartridge
3. Wipe the outside and flush the inside of the pick-up tubes with 70% ethanol. Let them dry
naturally.
Cleaning
Decalcification
Disinfection
When more than 7
days have passed
since last
disinfection
1. Run a heat disinfection program before treatment Disinfection
Every 1-3 months
if UFD is installed
1. Change the ultrafilter
2. Run a heat disinfection program
Disinfection