Preparing the machine - Concentrates
Turn the red concentrate connector clockwise and
then remove it from the standby port of the machine.
Connect the red concentrate connector to a pick up
tube until a ‘click’ is heard. Place the pick up tube into
the proper acidic concentrate container. Confirm the
concentrate combination.
Press the release
buttons. Open the upper
latch of the BiCart
cartridge holder. Pull it
outwards and upwards.
Open the lower latch.
Place the bottom port of
the BiCart cartridge into
the hole of the lower
latch.
Bring down the upper
latch to align the hole
with the top port of the
BiCart cartridge.
32
4/29/20