Preparing the machine attach the heparin syringe
4/29/20
35
It is important to correctly install the heparin syringe, it should be installed as
follows;
1. Insert the plastic collar of the syringe
into the groove of the pump
2. Insert the plate on the end of the
plunger into the groove on the piston
3. Turn the locking wheel upwards until
resistance is felt
4. Check the syringe is firmly in place by
lightly pulling on the plunger