Hemodialysis treatment end the treatment
When the treatment time is completed the operator’s is alerted by an audible
buzzer, a lit alarm indication light and an attention message.
After pressing confirm, the on screen instructions should be followed.
If the treatment needs to be finished before the set treatment time has expired,
then the time needs to be reduced down to 0:00 first. This is done by pressing the
time indicator and using the keypad to change the time.