Functions profiling ultrafiltration
4/29/20
71
With linear profiles the machine knows
the set time and UF volume, the operator
sets the start UF rate, and the machine
will calculate a smooth decreasing
profile.
With step profiles the machine knows the
set time and UF volume, the operator
sets the start UF rate, and the number of
steps and the machine will calculate the
profile.
With interval profiles the machine knows
the set time and UF volume, the operator
sets the start UF rate, and the number of
intervals and the machine will calculate
the profile.