Trouble shooting changing the dialyzer & blood lines
If the dialyzer and the blood lines need to be replaced during the treatment this
can be done by pressing the rinse back button.
In the rinse back menu there are two options; rinse back if the treatment should
be ended and the patient disconnected, or new blood circuit if the treatment
should be continued.
Within new blood circuit there are two options; rinse back if this is possible, or
new blood circuit if it is not possible or desired to rinse back.
The dialyzer fluid tubes are returned to the machine, and then the blood lines can
be removed. New blood lines and dialyzer are mounted. The new circuit is then
primed and then the patient can be reconnected.