Fibrin Sealant Surgery Spray set to be used Applicator tips to be used Pressure regulator to be used
Recommended distance
from target tissue
Recommended spray pressure/flow rate
TISSEEL
Open surgery
TISSEEL/ARTISS Spray Set n.a.
EASYSPRAY
10-15cm 1.5-2.0 bar (21.5-28.5 psi).
TISSEEL/ARTISS Spray Set 10 Pack n.a. EASYSPRAY
Laparoscopic/ minimally
invasive procedures
n.a.
DUPLOSPRAY MIS Applicator
20cm
DUPLOSPRAY MIS Regulator
2-5 cm
1.0-2.0 L/min
(1.2-1.5 bar/ 18-22 psi)
DUPLOSPRAY MIS Regulator NIST B11
DUPLOSPRAY MIS Applicator
30cm
DUPLOSPRAY MIS Regulator
DUPLOSPRAY MIS Regulator NIST B11
DUPLOSPRAY MIS Applicator
40cm
DUPLOSPRAY MIS Regulator
DUPLOSPRAY MIS Regulator NIST B11
Replaceable Tip
DUPLOSPRAY MIS Regulator
DUPLOSPRAY MIS Regulator NIST B11
ARTISS
Open surgery
TISSEEL/ARTISS Spray Set n.a.
EASYSPRAY
10-15cm 1.5-2.0 bar (21.5-28.5 psi)
TISSEEL/ARTISS Spray Set 10 Pack n.a. EASYSPRAY
Laparoscopic/ minimally
invasive procedures
n.a n.a n.a n.a n.a
CAUTION: Any application of pressurised gas may be associated with a potential risk of air or gas embolism, tissue rupture or air or gas entrapment with compression, which may be life threatening. Be sure to take
appropriate measures to address these risks by observing the recommended minimum spraying distance and the maximum pressure provided in the appropriate spray set instructions for use and in the table above.