12
Specifications (PW100 and PW200)
Class II
Type BF Applied Part (the collector tubing is the applied part)
IP21
Operating suction range: Approximately 120 mmHg – 140 mmHg (2.3 – 2.7 psi)
Maximum suction level: 182 mmHg (3.5 psi)
Electrical Specifications
AC input voltage: 100 –240VAC
AC frequency: 50 – 60Hz 10A
AC connector: NEMA 1-15P
Environmental Specifications
PW100 PW200
Operating
Temperature
Limit
41°F–104°F (5°C–40°C) 41°F–104°F (5°C–40°C)
Storage
Temperature Limit
-13°F–158°F (-25°C–70°C) -13°F–104°F (-25°C–40°C)
Transport
Temperature Limit
-13°F–158°F (-25°C–70°C) -13°F–158°F (-25°C–70°C)
Humidity
Limitation
5 –90%, non-condensing 5 –90%, non-condensing
Operating Atmospheric
Pressure Limitation
70 kPa - 106 kPa 70 kPa - 106 kPa
Storage/Transport
Atmospheric Pressure
Limitation
50 kPa - 106 kPa 50 kPa - 106 kPa
When the
PureWick™ Urine Collection System has been stored in a very cold
environment (down to -13°F or -25°C), please allow the device to sit for approximately
90 minutes at room temperature (approximately 68°F or 20°C) before use.
When the
PureWick™ Urine Collection System has been stored in a very warm
environment (up to 158°F or 70°C), please allow the device to sit for approximately 90
minutes at room temperature (approximately 68°F or 20°C) before use.
Physical Specifications
Dimensions with canister: 9” x 6.5” x 11” (22.86cm x 16.51cm x 27.94cm)
Dimensions without canister: 8.75” x 6” x 7.5” (22.22cm x 15.24cm x 19.05cm)
Cord length: 12’ (3.65m)
Battery Specifications (PW200 only)
Battery: Lithium-ion battery
Charging Time: 12 hours