E
ericksontomAug 20, 2025
How to fix a paper jam in BD FACSPresto printer?
- RRobert EvansAug 20, 2025
To fix a paper jam: 1. Open the printer door. 2. Clear the paper jam. 3. Close the printer door.
How to fix a paper jam in BD FACSPresto printer?
To fix a paper jam: 1. Open the printer door. 2. Clear the paper jam. 3. Close the printer door.
What to do if my BD Medical Equipment printer is out of paper?
If the printer is out of paper, simply add more printer paper.
What to do if BD Medical Equipment has low battery?
If the instrument indicates low battery power, plug it into a wall electrical outlet or another electric source. Then, press the power button to turn on the instrument.
What to do if the sample run has stopped on BD FACSPresto?
If a sample or process control run has stopped: 1. Wait 2 minutes. 2. If the progress bar does not move, press the power button until the instrument turns off. 3. Press the power button to turn on the instrument. A screen with the Open Door button appears. 4. Press Open Door. The cartridge is ejected. 5. Remove the cartridge.
Why does BD FACSPresto eject the cartridge during the run test?
The instrument might eject the cartridge during the run test because it cannot read the barcode or channel. To solve this: 1. Make sure the channel protector is removed from the cartridge. 2. Clean the barcode label or channel. 3. Re-insert the cartridge into the instrument. 4. If the error persists, prepare a new cartridge.
How to free up space on a full BD FACSPresto Medical Equipment USB flash drive?
To resolve a full USB flash drive, insert the USB flash drive into a PC and remove some files. Alternatively, insert another BD USB flash drive into the USB port.
What to do if the USB flash drive is not present in BD FACSPresto?
If the USB flash drive is not present, insert it into the USB port.
What to do if USB flash drive is not inserted correctly in BD FACSPresto?
If the USB flash drive is not inserted correctly, remove it and re-insert it into the USB port.
Why is my BD FACSPresto Medical Equipment shutting down after 35 minutes on battery?
If the instrument is running on battery and hasn't been used for 35 minutes, touch the screen to prevent shutdown. If it turns off, press the power button to turn it back on.
What to do if BD FACSPresto Medical Equipment temperature is out of range?
If the instrument temperature is outside its operational range: 1. Make sure the air vents are unobstructed. 2. Press the power button until the instrument turns off. 3. Wait for the instrument to return to its operational range. 4. Press the power button to turn on the instrument.
Laser Configuration | Blue (488 nm) |
---|---|
Software | BD FACSPresto Software |
Technology | Flow cytometry |
Sample Type | Whole blood |
Fluorochromes | PE, FITC |
Detectors | Photodiodes |
Power Requirements | 100–240 VAC, 50/60 Hz |
Data Storage | Internal storage, USB drive |
Explains the meaning of safety symbols and alerts used in the manual.
Guides users through the procedures for setting up, starting, and shutting down the instrument.
Details how to configure initial instrument settings like date, time, and display.
Provides steps for upgrading the instrument's software using a USB flash drive.
Covers running samples and inserting the cartridge into the instrument.
Offers solutions for common errors and problems encountered with the instrument.