Eltropol 300
4
Eltropol 300 is exclusively designed for the
electrolytic polishing of Co-Cr partial den-
ture bases using Wirolyt polishing solution.
Any other use is considered improper. We shall
not be liable for any damage resulting from im-
proper use.
One example of improper use is the heating of
water, other liquids, or foodstuffs!
The safety instructions must be followed in order
to avoid personal injury or damage to equipment.
We shall not be liable for failure to comply with
the safety instructions.
General safety instructions
l Place of use
The unit has been designed solely for use in
dental laboratories and comparable
establishments for research, commercial and
training purposes. • The unit may only be
installed on an adequately stable surface. • The
unit must only be used in dry rooms.
l Operator and operating personnel
The unit must only be operated by dental
professionals, or they must provide supervision if
trainees or other persons operate unit. • Wear
the protective clothing prescribed for use in
dental laboratories and observe the rules of
conduct. • The unit may only be operated by staff
members who are familiar with and follow the
instructions of this operating manual. • The
operating manual must be easily accessible when
the machine is operated. • Observe the national
accident prevention regulations in addition to
this operating manual.
l Handling the machine
Prior to connection to the power supply check
whether the specifications on the rating plate
correspond to the supply network. In the case of
uncertainty, consult an electrician. • The unit is
categorised under protection class 1 and may
only be connected to properly earthed power
sources (earth contact sockets). • Modifications
of the machine are not permitted. • Always keep
signs and stickers such that they are easy to read.
They must not be removed. • Check the machine
and leads for damage regularly. The machine
must not be operated if it has any defects that
may result in risks for the employees or third
persons. • Protect the leads against heat, oil and
sharp edges. The leads must not be used to carry
the machine or to pull out the mains plug.
l Care and maintenance
Always switch off the machine and pull out the
mains plug before performing care, cleaning or
maintenance work. • Only use dry or slightly
moist cloths for cleaning. Do not spray the
machine with water or immerse it in water.
l Repairs
Repairs may be carried out solely by BEGO
customer service or by persons authorised by
BEGO customer service. • Metallic housings
must be grounded properly so as to prevent them
from carrying current. Otherwise they are
extremely dangerous because the housings may
be live due to damage to the machine! Since a
check to ensure that the housing is in a de-
energized state is mandatory every time the
machine has been opened (Germany: check in
accordance with DIN VDE 0701-1), only
qualified electricians may open the machine! •
Only BEGO replacement and wearing parts may
be used.
Safety instructions