Eltropol 300
14
Changing the polishing solution
The more it is used, the less effective the polishing
solution becomes. The unit calculates and displays
the optimum point at which to renew the solution.
LED blinks: change solution soon.
LED lights up: change solution now.
CAUTION
Wirolyt contains glycol and sulphuric acid!
Avoid contact with skin and eyes!
Risk of blinding!
Wear safety glasses and safety gloves!
Never add water!
Dispose of as hazardous waste! Used Wirolyt
contains heavy metal ions – do not allow it
enter the sewer system!
Switch off the unit at the main switch and
pull the plug out.
Carefully turn the unit around without tipping it in
order to reach the drainage tube on the back of
the unit.
Pull the drainage tube out of the clamp, hold it
upright and remove the stopper. Lower the tube
into a suitable canister with a capacity of at least
two litres and allow the polishing solution to drain
off. Use a funnel if necessary.
Risk of scalding!
The operating temperature of Wirolyt can
reach 40-60°C!
Start
+