Installation
• The unit must be placed on a sufficiently stable
surface.
• The unit may only be used in an enclosed space.
Prior to making the electrical
connection, check whether the voltage
selector switch (on the rear side) is in
the correct position for the mains
power supply. In the event of
uncertainty consult a specialised staff
member.
• The unit is categorised under protection class 1
and may only be connected to properly earthed
power sources (earth contact sockets).
• Use the appropriate power cable (EU or USA plug)
included with the unit.
Note
The voltage selector switch is protected in a recess
approx. 5 cm inside the back wall of the unit and only
visible when sufficient lighting is available. Ex works it
is set at 230 V (see adhesive label).
The two 3.15 A fuses secure the entire voltage range
from 90 V to 260 V.