Check seals
After the O-ring seals are cleaned, you can check
whether compressed air still leaks:
1. Select menu Settings / O-ring seals / Check
seals.
2. During the check the chambers are filled with
compressed air. If no air escapes (generation
of noise!), the O-ring seals function properly.
Drain the pressure reducers
The two pressure reducers must be drained after
every 500 castings.
1. Switch off the machine and pull out the mains
plug.
2. Remove the panel (1) on the left side (loosen
knurled screws).
3. Have a container or a cloth at hand to collect
the water.
4. To drain the water, turn the lower screw (2, 3)
counter-clockwise all the way to the stop.