Table 2
For all me equipment and me systems
Guidance and manufacturers declaration electromagnetic immunity
Guidance and manufacturer’s declaration electromagnetic immunity
BM55
is intended for use in the electromagnetic environment specified below. The customer or the user of
BM55
should assure that it is
used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or cera-
mic tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30 %.
Electrical fast transient/burst
IEC 61000-4-4
± 2 kV for power supply lines
± 1 kV input/output lines
± 2 kV for power supply lines
± 1 kV input/output lines
Mains power quality should be that of a typi-
cal commercial or hospital environment.
Surge IEC 61000-4-5
± 1 kV dierential mode
± 2 kV common mode
± 1 kV dierential mode
± 2 kV common mode
Mains power quality should be that of a typi-
cal commercial or hospital environment.
Voltage dips, short interrup-
tions and voltage variations
on power supply input lines
IEC 61000-4-11
< 5 % U
T
(> 95 % dip in U
T
)
for 0,5 cycle
40 % U
T
(60 % dip in U
T
)
for 5 cycles
70 % U
T
(30 % dip in U
T
)
for 25 cycles
< 5 % U
T
(> 95 % dip in U
T
)
for 5 sec
< 5 % U
T
(> 95 % dip in U
T
)
for 0,5 cycle
40 % U
T
(60 % dip in U
T
)
for 5 cycles
70 % U
T
(30 % dip in U
T
)
for 25 cycles
< 5 % U
T
(> 95 % dip in U
T
)
for 5 sec
Mains power quality should be that of a typi-
cal commercial or hospital environment. If
the user of
BM55
requires continued ope-
ration during power mains interruptions, it
is recommended that
BM55
is be powered
from an uninterruptible power supply or a
battery.
Power frequency (50/ 60 Hz)
magnetic field IEC 61000-4-8
3 A/m 3 A/m
Power frequency magnetic fields should be
at levels characteristic of a typical location
in a typical commercial or hospital envi-
ronment.
NOTE: U
T
is the a.c. mains voltage prior to application of the test level.