Table 3
For all me equipment and me systems that are not life-supporting
Guidance and manufacturers declaration electromagnetic immunity
Guidance and manufacturer’s declaration electromagnetic immunity
BM55
is intended for use in the electromagnetic environment specified below. The customer or the user of
BM55
should assure that it is
used in such an environment.
Immunity test IEC 60601 test
level
Compli-
ance level
Electromagnetic environment – guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to
80 MHz
3 V/m
80 MHz to
2,5 GHz
3 V
3 V/ m
Portable and mobile RF communications equipment should be used no closer to any
part of
BM55
, including cables, than the recommended separation distance calculated
from the equation applicable to the frequency of the transmitter.
Recommended separation distance:
d = 1,2
d = 1,2 80 MHz to 800 MHz
d = 2,3 800 MHz to 2,5 GHz
Where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer and d is the recommended separation distance in metres
(m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site
survey,
a
should be less than the compliance level in each frequency range.
b
Interference may occur in the vicinity of equipment marked with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is aected by absorption and reflection from
structures, objects and people.