KBF (E5.2)
Constant climate chambers with program control
Service Manual
valid for the following models of KBF, starting with serial no. 09-11327:
Model
order no.
KBF 115 (E5.2)
9020-0139, 9120-0139
KBF 115-UL (E5.2)
9020-0184, 9120-0184
KBF 240 (E5.2)
9020-0150, 9120-0150
KBF 240-UL (E5.2)
9020-0176, 9120-0176
KBF 720 (E5.2)
9020-0108, 9120-0108
KBF 720-UL (E5.2)
9020-0168, 9120-0168
Issue 04-2015 order no. 7001-0203

Table of Contents

Other manuals for Binder KBF 115

Questions and Answers

  • C
    Charles AllenAug 19, 2025
    What to do if the Binder Laboratory Equipment chamber doesn’t heat up and the safety controller responds?
    • D
      Donald BeckerSep 12, 2025
      If the Binder Laboratory Equipment chamber is not heating up when turned on and the safety controller is responding, first acknowledge the alarm on the controller. Then, check the temperature setpoint setting. If necessary, select a suitable safety controller setpoint.
  • L
    lisadayAug 21, 2025
    How to fix low refrigerating performance in Binder Laboratory Equipment?
    • M
      morganrachaelAug 21, 2025
      To address low or no refrigerating performance in your Binder Laboratory Equipment, ensure the ambient temperature is not higher than 25 °C / 77 °F by selecting a cooler place of installation. Also, make sure that the combination of temperature and humidity values are within the optimum range. Reduce any excessive external heat load. For compressor not turned on case, contact BINDER service.
  • K
    Kayla WilsonAug 23, 2025
    Why is my Binder Laboratory Equipment chamber heating permanently?
    • L
      Leslie BondAug 23, 2025
      If the Binder Laboratory Equipment chamber is heating permanently and not maintaining the set-point, calibrate and adjust the controller. For controller defective and semiconductor relay defective cases, contact BINDER service.
  • D
    dakotabarrettAug 26, 2025
    How to resolve humidity fluctuation in Binder Laboratory Equipment?
    • M
      michaeljordanAug 26, 2025
      To resolve humidity fluctuations where the control accuracy of ± 3 % r.F. is not reached in your Binder Laboratory Equipment, replace the door gasket if it's defective. Additionally, reduce how frequently the door is opened.
  • D
    Daniel KennedyAug 27, 2025
    What to do if the mechanical safety device class 3.2 responds in Binder Laboratory Equipment?
    • C
      caitlin14Aug 27, 2025
      If the mechanical safety device class 3.2 responds in your Binder Laboratory Equipment (with option safety device class 3.3), acknowledge the alarm on the controller. Check the temperature set-point and safety device class 3.2 settings. Adjust the limit value as needed. For controller or safety device defective case, contact BINDER service.
  • C
    Christopher RichardsonAug 29, 2025
    What to do if there is an electrical hazard during live maintenance work on a Binder climate chamber?
    • M
      michaelcarpenterAug 30, 2025
      To avoid electrical hazards during maintenance, ensure the Binder climate chamber does not get wet. Do not remove the rear panel and always disconnect the chamber by turning off the main power switch and pulling the power plug before starting maintenance. Ensure that only licensed electricians or experts authorized by BINDER conduct general maintenance. Maintenance on the refrigeration system should only be performed by qualified personnel trained according to EN 13313:2010, such as a refrigeration technician with certified expert knowledge, following national statutory regulations.
  • G
    Glenda RiveraSep 1, 2025
    How to troubleshoot low dehumidification in Binder Laboratory Equipment?
    • E
      Eric CharlesSep 1, 2025
      To address low or no dehumidification in your Binder Laboratory Equipment, turn on the humidity control.
  • K
    Kathleen WebbSep 3, 2025
    How to deal with humidity and temperature fluctuation in Binder Laboratory Equipment?
    • A
      awilsonSep 3, 2025
      To correct humidity and temperature fluctuations greater than 1 °C with a set-point approximately 3 °C above ambient temperature in your Binder Laboratory Equipment, select a cooler place of installation or contact BINDER service.
  • F
    Frank ReynoldsSep 5, 2025
    How to prevent condensation in the inner chamber of Binder Laboratory Equipment?
    • C
      Caroline GuerreroSep 5, 2025
      If you see condensation at the walls of the inner chamber of your Binder Laboratory Equipment, select a combination of temperature and humidity values in the optimum range. If the set-point was too long below ambient temperature, defrost the chamber.
  • P
    Patricia GallegosSep 7, 2025
    How to improve humidity and temperature accuracy in Binder Laboratory Equipment?
    • M
      mkaneSep 7, 2025
      To improve low humidity and temperature accuracy in your Binder Laboratory Equipment, set the fan speed to 100%.

Summary

Preface

Safety Instructions Structure

Explains the structure and levels of safety instructions, warning symbols, and imperative symbols used.

General Safety Instructions

Qualification of Service Personnel

Specifies the necessary qualifications and training for personnel performing maintenance and repair.

Dangers and Hazards

Details potential dangers like electric shock, cutting injuries, mechanical damages, and water hazards.

Maintenance

Safety Check Procedures

Details safety checks according to DIN VDE 0701-0702, including visual inspection and resistance measurements.

Troubleshooting

General Electrical Faults

Provides a troubleshooting guide for common electrical faults, including power and circuit breaker issues.

Heating System Faults

Guides through diagnosing and resolving issues related to the heating system components.

Cooling System Faults

Offers troubleshooting steps for problems occurring in the cooling system.

Humidification Module Errors

Details error localization and causes for malfunctions in the humidification module.

Repair Work at the Housing

Repair Work at the Outer Door

Repair Work at the Inner and Outer Chamber

Repair Work at Electric and Electronic Components

Humidifying and Dehumidifying System

Repair Work at the Cooling System

General Information and Safety

Provides warnings, cautions, and qualifications required for cooling system repair.

Calibrating and Adjusting Temperature and Humidity Control

General Execution of Calibration

Outlines the general procedure for calibration, including instrument setup and steady-state conditions.

Calibrating Temperature Control

Details on setting calibration points and performing adjustments for the temperature controller.

Calibrating Humidity Control

Steps for calibrating and adjusting the humidity controller at two set points.

Binder KBF 115 Specifications