EasyManua.ls Logo

Binder KBF 720 - Service Manual

Default Icon
140 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
KBF (E5.2)
Constant climate chambers with program control
Service Manual
valid for the following models of KBF, starting with serial no. 09-11327:
Model
order no.
KBF 115 (E5.2)
9020-0139, 9120-0139
KBF 115-UL (E5.2)
9020-0184, 9120-0184
KBF 240 (E5.2)
9020-0150, 9120-0150
KBF 240-UL (E5.2)
9020-0176, 9120-0176
KBF 720 (E5.2)
9020-0108, 9120-0108
KBF 720-UL (E5.2)
9020-0168, 9120-0168
Issue 04-2015 order no. 7001-0203

Table of Contents

Other manuals for Binder KBF 720

Questions and Answers

  • J
    Jonathan JacksonAug 19, 2025
    What to do if the Binder Laboratory Equipment chamber doesn’t heat up and the safety controller responds?
    • A
      Angela MichaelSep 12, 2025
      If the Binder Laboratory Equipment chamber is not heating up when turned on and the safety controller is responding, first acknowledge the alarm on the controller. Then, check the temperature setpoint setting. If necessary, select a suitable safety controller setpoint.
  • N
    Nathan FosterAug 21, 2025
    How to fix low refrigerating performance in Binder Laboratory Equipment?
    • K
      Kristen BeltranAug 21, 2025
      To address low or no refrigerating performance in your Binder Laboratory Equipment, ensure the ambient temperature is not higher than 25 °C / 77 °F by selecting a cooler place of installation. Also, make sure that the combination of temperature and humidity values are within the optimum range. Reduce any excessive external heat load. For compressor not turned on case, contact BINDER service.
  • J
    Jennifer RiversAug 23, 2025
    Why is my Binder Laboratory Equipment chamber heating permanently?
    • S
      Sheila GregoryAug 23, 2025
      If the Binder Laboratory Equipment chamber is heating permanently and not maintaining the set-point, calibrate and adjust the controller. For controller defective and semiconductor relay defective cases, contact BINDER service.
  • P
    Patrick MitchellAug 26, 2025
    How to resolve humidity fluctuation in Binder Laboratory Equipment?
    • M
      Miguel RobersonAug 26, 2025
      To resolve humidity fluctuations where the control accuracy of ± 3 % r.F. is not reached in your Binder Laboratory Equipment, replace the door gasket if it's defective. Additionally, reduce how frequently the door is opened.
  • A
    acevedoedgarAug 27, 2025
    What to do if the mechanical safety device class 3.2 responds in Binder Laboratory Equipment?
    • J
      Joseph ClarkAug 28, 2025
      If the mechanical safety device class 3.2 responds in your Binder Laboratory Equipment (with option safety device class 3.3), acknowledge the alarm on the controller. Check the temperature set-point and safety device class 3.2 settings. Adjust the limit value as needed. For controller or safety device defective case, contact BINDER service.
  • P
    Paul RodriguezAug 30, 2025
    What to do if there is an electrical hazard during live maintenance work on a Binder KBF 720 Climate chamber?
    • C
      Curtis StoutAug 30, 2025
      To avoid electrical hazards during maintenance, ensure the Binder climate chamber does not get wet. Do not remove the rear panel and always disconnect the chamber by turning off the main power switch and pulling the power plug before starting maintenance. Ensure that only licensed electricians or experts authorized by BINDER conduct general maintenance. Maintenance on the refrigeration system should only be performed by qualified personnel trained according to EN 13313:2010, such as a refrigeration technician with certified expert knowledge, following national statutory regulations.
  • S
    Susan GonzalezAug 31, 2025
    How to troubleshoot low dehumidification in Binder Laboratory Equipment?
    • J
      Jeremy LeeSep 1, 2025
      To address low or no dehumidification in your Binder Laboratory Equipment, turn on the humidity control.
  • J
    Jeffrey GreenSep 3, 2025
    How to deal with humidity and temperature fluctuation in Binder Laboratory Equipment?
    • J
      Joseph MacdonaldSep 3, 2025
      To correct humidity and temperature fluctuations greater than 1 °C with a set-point approximately 3 °C above ambient temperature in your Binder Laboratory Equipment, select a cooler place of installation or contact BINDER service.
  • K
    Karl OconnorSep 5, 2025
    How to prevent condensation in the inner chamber of Binder Laboratory Equipment?
    • E
      Ernest HunterSep 5, 2025
      If you see condensation at the walls of the inner chamber of your Binder Laboratory Equipment, select a combination of temperature and humidity values in the optimum range. If the set-point was too long below ambient temperature, defrost the chamber.
  • J
    Jessica CamachoSep 7, 2025
    How to improve humidity and temperature accuracy in Binder Laboratory Equipment?
    • B
      brighttimothySep 7, 2025
      To improve low humidity and temperature accuracy in your Binder Laboratory Equipment, set the fan speed to 100%.

Summary

General Safety Instructions

Personnel Qualification, Hazards, and Electrical Connection

Details service personnel requirements, potential dangers, and electrical connection specifications.

Maintenance

Troubleshooting

General Electrical Faults

Guides on diagnosing and resolving common electrical issues within the unit.

Heating System Faults

Provides troubleshooting steps for issues related to the unit's heating system.

Cooling System Faults

Offers solutions for identifying and fixing problems within the cooling system.

Humidification Module Errors

Details troubleshooting procedures for various errors in the humidification system.

Repair Work at the Inner and Outer Chamber

Repair Work at the Cooling System

General Information and Tools

Provides safety warnings, required qualifications, and special tools for cooling system repairs.

Calibrating and Adjusting Control Systems

General Calibration and Temperature Control

Covers general calibration steps, setting points, door heating correction, and adjusting the temperature controller.

Humidity Control Calibration and Adjustment

Details calibrating and adjusting the humidity controller, including reset and setting points.

Binder KBF 720 Specifications

General IconGeneral
Temperature fluctuation±0.3 °C
Humidity range10% to 98% RH
Humidity fluctuation±1.5% RH
Voltage230 V
Frequency50/60 Hz
Interior volume700 liters

Related product manuals