KBF (E5.2)
Constant climate chambers with program control
Service Manual
valid for the following models of KBF, starting with serial no. 09-11327:
Model
order no.
KBF 115 (E5.2)
9020-0139, 9120-0139
KBF 115-UL (E5.2)
9020-0184, 9120-0184
KBF 240 (E5.2)
9020-0150, 9120-0150
KBF 240-UL (E5.2)
9020-0176, 9120-0176
KBF 720 (E5.2)
9020-0108, 9120-0108
KBF 720-UL (E5.2)
9020-0168, 9120-0168
Issue 04-2015 order no. 7001-0203

Table of Contents

Other manuals for Binder KBF 240

Questions and Answers

  • J
    Jason FullerAug 21, 2025
    How to fix low refrigerating performance in Binder Laboratory Equipment?
    • A
      Anthony GreeneAug 21, 2025
      To address low or no refrigerating performance in your Binder Laboratory Equipment, ensure the ambient temperature is not higher than 25 °C / 77 °F by selecting a cooler place of installation. Also, make sure that the combination of temperature and humidity values are within the optimum range. Reduce any excessive external heat load. For compressor not turned on case, contact BINDER service.
  • C
    christine71Aug 23, 2025
    Why is my Binder Laboratory Equipment chamber heating permanently?
    • A
      Angela CarpenterAug 23, 2025
      If the Binder Laboratory Equipment chamber is heating permanently and not maintaining the set-point, calibrate and adjust the controller. For controller defective and semiconductor relay defective cases, contact BINDER service.
  • D
    Darlene MathisAug 26, 2025
    How to resolve humidity fluctuation in Binder Laboratory Equipment?
    • T
      Tiffany SimmonsAug 26, 2025
      To resolve humidity fluctuations where the control accuracy of ± 3 % r.F. is not reached in your Binder Laboratory Equipment, replace the door gasket if it's defective. Additionally, reduce how frequently the door is opened.
  • J
    Jesse GreeneAug 27, 2025
    What to do if the mechanical safety device class 3.2 responds in Binder Laboratory Equipment?
    • C
      Christine WilliamsAug 28, 2025
      If the mechanical safety device class 3.2 responds in your Binder Laboratory Equipment (with option safety device class 3.3), acknowledge the alarm on the controller. Check the temperature set-point and safety device class 3.2 settings. Adjust the limit value as needed. For controller or safety device defective case, contact BINDER service.
  • M
    Michael MathisAug 30, 2025
    What to do if there is an electrical hazard during live maintenance work on a Binder KBF 240 Climate chamber?
    • O
      oguerreroAug 30, 2025
      To avoid electrical hazards during maintenance, ensure the Binder climate chamber does not get wet. Do not remove the rear panel and always disconnect the chamber by turning off the main power switch and pulling the power plug before starting maintenance. Ensure that only licensed electricians or experts authorized by BINDER conduct general maintenance. Maintenance on the refrigeration system should only be performed by qualified personnel trained according to EN 13313:2010, such as a refrigeration technician with certified expert knowledge, following national statutory regulations.
  • N
    Nancy BurnsSep 1, 2025
    How to troubleshoot low dehumidification in Binder Laboratory Equipment?
    • B
      Brittany CarrollSep 1, 2025
      To address low or no dehumidification in your Binder Laboratory Equipment, turn on the humidity control.
  • R
    Rebecca MillerSep 3, 2025
    How to deal with humidity and temperature fluctuation in Binder Laboratory Equipment?
    • E
      Elizabeth EspinozaSep 3, 2025
      To correct humidity and temperature fluctuations greater than 1 °C with a set-point approximately 3 °C above ambient temperature in your Binder Laboratory Equipment, select a cooler place of installation or contact BINDER service.
  • R
    richard12Sep 5, 2025
    How to prevent condensation in the inner chamber of Binder Laboratory Equipment?
    • K
      Kevin SteeleSep 5, 2025
      If you see condensation at the walls of the inner chamber of your Binder Laboratory Equipment, select a combination of temperature and humidity values in the optimum range. If the set-point was too long below ambient temperature, defrost the chamber.
  • R
    Raymond LozanoSep 7, 2025
    How to improve humidity and temperature accuracy in Binder Laboratory Equipment?
    • J
      Jessica BeckSep 7, 2025
      To improve low humidity and temperature accuracy in your Binder Laboratory Equipment, set the fan speed to 100%.
  • M
    melindamillerSep 9, 2025
    What to do if my Binder Laboratory Equipment chamber has no function and a dark display?
    • A
      Angel OwensSep 9, 2025
      If there is no chamber function and the display is dark on your Binder Laboratory Equipment, try pressing on the touchscreen. Also, make sure that the main power switch is turned on.

Summary

General safety instructions

Qualification of service personnel

Requirements for technicians performing maintenance and repairs.

Dangers

Identification of potential hazards (electric shock, mechanical, etc.).

Maintenance

Maintenance schedule for the annual maintenance work

Overview of tasks to be performed during annual service.

Safety check acc. to DIN VDE 0701-0702

Electrical safety testing procedures after repair or modification.

Troubleshooting

General electrical faults

Common electrical problems and their solutions.

Faults of the heating system

Troubleshooting steps for heating system malfunctions.

Faults of the cooling system

Troubleshooting steps for cooling system malfunctions.

Errors in the humidification module

Diagnosis and correction of humidification module faults.

Repair work at the housing

Repair work at the outer door

Checking and replacing outer door components

Guidance on inspecting and replacing parts within the outer door.

Repair work at the inner and outer chamber

Repair work at electric and electronic components

Humidifying and dehumidifying system

Repair work at the cooling system

Calibrating and adjusting the temperature and humidity control