Operating Manual
Translation of the original operating manual
KBF P (E6)
Constant climate chambers with program control
and adjustable light cassettes
Model Model version Art. No.
KBF P 240 KBFP240-230V 9020-0328, 9120-0328
KBF P 240-UL KBFP240UL-240V 9020-0329, 9120-0329
KBF P 720 KBFP720-230V 9020-0330, 9120-0330
KBF P 720-UL KBFP720UL-240V 9020-0331, 9120-0331
KBF LQC (E6)
Constant climate chambers
with ICH compliant illumination and light dose detection
with program control and adjustable light cassettes
Model Model version Art. No.
KBF LQC 240 KBFLQC240-230V 9020-0332, 9120-0332
KBF LQC 240-UL KBFLQC240UL-240V 9020-0333, 9120-0333
KBF LQC 720 KBFLQC720-230V 9020-0334, 9120-0334
KBF LQC 720-UL KBFLQC720UL-240V 9020-0335, 9120-0335
KBWF (E6)
Growth chambers with light and humidity
with program control and adjustable light cassettes
Model Model version Art. No.
KBWF 240 KBWF240-230V 9020-0336, 9120-0336
KBWF 720 KBWF720-230V 9020-0337, 9120-0337
BINDER GmbH
Address: Post office box 102, 78502 Tuttlingen, Germany Phone: +49 7462 2005 0
Fax: +49 7462 2005 100 Internet: http://www.binder-world.com
E-mail: info@binder-world.com Service Hotline: +49 7462 2005 555
Service Fax: +49 7462 2005 93 555 Service E-Mail: service@binder-world.com
Service Hotline USA: +1 866 885 9794 or +1 631 224 4340 x3
Service Hotline Asia Pacific: +852 390 705 04 or +852 390 705 03
Service Hotline Russia and CIS: +7 495 988 15 16

Table of Contents

Questions and Answers

  • Z
    Zachary SmithAug 11, 2025
    What to do if my Binder KBF P 240 Temperature Controller chamber has no function?
    • A
      Ann OconnellAug 12, 2025
      If the chamber isn't working, first check the power supply connection. Ensure the voltage is correct. Inspect the chamber fuse and replace it if needed. If the fuse blows again, contact BINDER service. Also, the over temperature protective device (class 1) might respond if nominal temperature exceeded by 10° due to chamber failure.
  • B
    blesterAug 18, 2025
    Why is my Binder KBF P 240 Temperature Controller chamber heating permanently and not maintaining the set-point?
    • R
      Randy HughesAug 18, 2025
      If your chamber is constantly heating and not holding the set-point, the controller may be defective, in which case you should contact BINDER service. Alternatively, the controller may need calibration and adjustment. Also, semiconductor relay might be defective.
  • D
    David McclureAug 26, 2025
    What to do if my Binder Temperature Controller chamber doesn’t heat up when turned on and the safety controller responds?
    • N
      Nina OrtizAug 26, 2025
      If the chamber doesn't heat up and the safety controller is active, acknowledge the alarm on the controller. Check the temperature setpoint setting. The inner chamber temperature might have reached the safety controller set-point, or the safety controller set too low. If appropriate, select suitable safety controller setpoint.
  • M
    Mr. Zachary SingletonSep 2, 2025
    Why doesn’t my Binder Temperature Controller chamber heat up?
    • K
      Kristine AllenSep 3, 2025
      If the chamber isn't heating up, the Pt 100 sensor might be defective, so contact BINDER service. The heating element or semiconductor relay might also be defective.

Binder KBF P 240 Specifications

General IconGeneral
BrandBinder
ModelKBF P 240
CategoryTemperature Controller
LanguageEnglish