KBF P + KBF LQC + KBWF (E6) 11/2016 page 164/189
24. Troubleshooting
Chamber without function.
No power supply.
Check connection to power sup-
ply.
Wrong voltage.
Check power supply for correct
voltage (chap. 4.6).
Chamber fuse has responded.
Check chamber fuse and replace
it if appropriate. If it responds
again, contact BINDER service.
Contact BINDER service.
Nominal temperature exceeded
by 10° due to chamber failure.
Over temperature protective de-
Chamber heating permanently,
set-point not maintained.
Contact BINDER service.
Semiconductor relay defective.
Controller not well adjusted, or
adjustment interval exceeded.
Calibrate and adjust controller.
Chamber doesn’t heat up.
Contact BINDER service.
Heating element defective.
Semiconductor relay defective
Chamber doesn’t heat up when
turned on.
Safety controller responds.
Inner chamber temperature has
reached the safety controller set-
point. Safety controller set too
low.
Acknowledge the alarm on the
controller. Check temperature
setpoint setting. If appropriate,
select suitable safety controller
Safety controller (chap. 13.2)
defective.
Contact BINDER service.
Mechanical safety device class
3.1 responds
(with option safety device class
3.3).
Limit temperature reached.
Acknowledge the alarm on the
controller. Check setting of tem-
perature set-point and safety
device class 3.1. If appropriate,
select suitable limit value.
Too much external heat load.
Contact BINDER service.
Semi-conductor relay defective
Mechanical safety device class
3.2 responds
(with option safety device class
3.3).
Limit temperature reached.
Acknowledge the alarm on the
controller. Check setting of tem-
perature set-point and safety
device class 3.2. If appropriate,
select suitable limit value.
Controller or safety device defec-
tive.
Contact BINDER service.