EasyManua.ls Logo

Binder KBF P 240 - Alarm Messages; Messages Concerning the Humidity System

Binder KBF P 240
189 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
KBF P + KBF LQC + KBWF (E6) 11/2016 page 93/189
12.1.3 Alarm messages
Condition Alarm message
Start after
condition occurred
Zero-voltage relay
alarm output (option)
The current actual temperature value
is outside the tolerance range (chap.
12.4)
Temperature range
after configurable
time
time as alarm start
The current actual humidity value is
outside the tolerance range (chap.
12.4)
Humidity range
after configurable
time
time as alarm start
Chamber door open
Door open
after 5 minutes ----
Power failure --- ---- immediately
Exceeded setpoint of the safety control-
ler class 3.1
Safety controller
immediately ----
Exceeded maximum or minimum tem-
perature (option temperature safety
device class 3.3)
Temp. safety de-
vice
immediately ----
Temperature sensor defective
e.g. - - - - or
<-<-< or >->->
immediately ----
Safety controller temperature sensor
defective
Safety controller
sensor
immediately ----
KBF LQC: Both dose target values VIS
and UVA reached
VIS and UVA dos-
es reached
immediately immediately
At least one light sensor plugged in:
Maximum temperature limited to 60 °C
Light sensor 60 °C.
immediately ----
Alarm messages are displayed in the list of active alarms until acknowledging them. They are also shown
in the event list.
12.1.4 Messages concerning the humidity system
Condition and measures Message
Start after condition
occurred
The humidity module is defective. Contact BINDER ser-
vice
Humidity system
immediately
The humidity module cannot fill up.
In case of freshwater supply via water pipe:
The water tap is closed, or the chamber is defective.
In case of freshwater supply via freshwater can (option,
chap. 21.9):
Water can is empty. Humidification is turned off. In case
of refrigerating operation, the interior is strongly dehu-
midified. When the water supply is functional again, the
humidity system restarts, or the chamber is defective.
Freshwater supply
immediately
The humidity module cannot empty the condensate
tank.
Check the length and location of the wastewater tube. If
appropriate contact BINDER service.
Waste water
immediately
Purging is required soon. Turn off and on again the hu-
midity switch to start purging. After a successful purging
the notification resets automatically.
Humidity system
purging req.
after predefined time
(approx. 3-5 months
depending on use)

Table of Contents

Related product manuals