5. Functional overview of the MB2 chamber controller
The MB2 chamber controller controls following parameters inside the chamber: temperature and humidity
(closed loop control), and fan speed.
Channel 1: Temperature in °C
Channel 2: Relative humidity in % r.H.
Illumination is controlled via the operation lines.
KBF LQC: In addition to displaying the actual values of UVA and the visible spectral range, the function
“Light Quantum Control” permits cumulative measurement of the light doses. In Manual Mode you can
enter target dose values of UVA and the visible spectral range. When they are reached, the UVA, and cool
white fluorescent tubes automatically turn off independently from each other.
For the control ranges of temperature and humidity, see climatic diagrams (chap. 18).
You can enter the desired set point values in fixed value operation mode directly on the display surface or
via the setpoint menu. For program operation the controller offers programming week and time programs.
In addition there is a timer program available (stopwatch function).
The controller offers various notifications and alarm messages with visual and audible indication and re-
mote alarms via e-mail, an event list (trace file) and the graphical display of the measuring values in the in
der chart recorder view. The MB2 program controller permits programming temperature and humidity
cycles. You can enter values or programs directly at the controller or use the APT-COM™ 3 DataControl-
System software (option) specially developed by BINDER.