KBF P + KBF LQC + KBWF (E6) 11/2016 page 94/189
Condition and measures Message
Start after condition
occurred
Maintenance of the humidity system is required. Contact
BINDER service.
Humidity module
service
after predefined time
(approx. 1 year)
Messages concerning the humidity system are shown in the event list.
When operating the chamber without water connection, turn off humidity control in the “set-
points” menu (chap. 6.3) in order to avoid humidity alarms.
12.2 State of alarm
1. Visual indications in Normal display: alarm message, screen header flashing in red color
2. Audible alert, if the buzzer is enabled (chap. 12.5).
3. Switching the zero-voltage relay alarm output (option, chap. 21.5) to transmit the alarm e.g., to a cen-
tral monitoring system.
Normal display in state of alarm (example).
(a) S
creen header flashing in red color and
showing the alarm message
(b) Alarm icon on the bottom of the screen:
change to the list of active alarms and alarm
acknowlegdement
(c) If applicable, information icon in the screen
header. Indication of a certain condition
12.3 Resetting an alarm, list of active alarms
Normal display in state of alarm (example).
Press the Alarm icon
List of active alarms.
Press the Reset alarm icon.
(a)
(c)
(b)