KBF P + KBF LQC + KBWF (E6) 11/2016 page 94/189
Condition and measures Message
Start after condition
occurred
Maintenance of the humidity system is required. Contact
BINDER service.
Humidity module
service
after predefined time
(approx. 1 year)
Messages concerning the humidity system are shown in the event list.
When operating the chamber without water connection, turn off humidity control in the “set-
points” menu (chap. 6.3) in order to avoid humidity alarms.
12.2 State of alarm
1. Visual indications in Normal display: alarm message, screen header flashing in red color
2. Audible alert, if the buzzer is enabled (chap. 12.5).
3. Switching the zero-voltage relay alarm output (option, chap. 21.5) to transmit the alarm e.g., to a cen-
tral monitoring system.
Normal display in state of alarm (example).
(a) S
creen header flashing in red color and
showing the alarm message
(b) Alarm icon on the bottom of the screen:
change to the list of active alarms and alarm
acknowlegdement
(c) If applicable, information icon in the screen
header. Indication of a certain condition
12.3 Resetting an alarm, list of active alarms
Normal display in state of alarm (example).
Press the Alarm icon
List of active alarms.
Press the Reset alarm icon.