KBF P + KBF LQC + KBWF (E6) 11/2016 page 55/189
8.3 Automatic correction of the actual value when turning on or off the
illumination
The chambers have been adjusted for operation with maximum illumination. Since the illumination creates
a heat input in the chamber, this is considered automatically when operating without illumination.
This can be recognized when turning on or off the illumination by a change of the actual temperature and
humidity values, which subsequently will equilibrate again to the set-points.
8.4 Light commutation and special controller functions via operation lines
You can define the switching state of up to 16 operation lines (control contacts). They are used to activate
/ deactivate special controller functions.
• Operation line “Humidity off” serves to turn off the humidity.
• Operation line “Idle mode” activates / deactivates the operating mode “Idle mode” (chap. 5.4).
• KBF P / KBF LQC: Operation line “Light VIS” serves to turn on/off the cool white fluorescent tubes
• KBF P / KBF LQC: Operation line “Light UVA” serves to turn on/off the BINDER Synergy Light fluores-
cent tubes
• KBWF: Operation lines “Light level 1” and “Light level 2” serve to turn on/off the fluorescent tubes
• KBF LQC: Operation line “LQC An” serves to turn on/off the light integration function.
• KBF LQC: Operation line “LQC reset VIS” serves to reset to zero the integrated VIS values once in a
time.
• KBF LQC: Operation line “LQC reset UVA” serves to reset to zero the integrated UVA values once in a
time.
The other operation lines are without function.