Refrigerating performance
Low or no refrigerating perfor-
mance.
Ambient temperature > 25 °C / 77
Select cooler place of installation.
ture/humidity values not in the
optimum range (see temperature
humidity diagram, chap. 18).
Select combination of tempera-
ture/humidity values in the opti-
mum range (chap. 18).
Compressor not turned on.
Contact BINDER service.
Electro-valves defective.
No or not enough refrigerant.
Too much external heat load.
Control accuracy of +/- 3 % r.h. is
Door opened very frequently. Open doors less frequently.
Humidity fluctuation, together with
temperature fluctuation >
1 °C with a set-point ca. 3 °C
above ambient temperature.
Place of installation too hot.
Select cooler place of installation
or contact BINDER service.
Low or no dehumidification.
Contact BINDER service.
Humidity control turned off.
Turn on humidity control (chap.
Icing at the evaporator plates.
Set-point was too long below
ambient temperature.
Defrost the chamber (chap. 19).
Condensation at the walls of the
inner chamber.
ture/humidity values not in the
optimum range (see temperature
humidity diagram, chap. 18)
Select combination of tempera-
ture/humidity values in the opti-
mum range (chap. 18).
Set-point was too long below
ambient temperature, icing in the
Defrost the chamber (chap. 19)
Low humidity and temperature
Fan speed has been reduced. Set fan speed to 100%.
No chamber function
(dark display).
Main power switch turned off.
Turn on the main power switch.
Menu functions not available.
Menu functions not available with
current authorization level.
Log in with the required higher
authorization.
Chart recorder function: meas-
ured-value memory cleared; in-
New setting of storage rate or
scaling (minimum and/or maxi-
Change the storage rate or scal-
ing ONLY if the previously regis-
tered data are no longer required.
Controller does not equilibrate to
setpoints entered in Fixed value
Controller is not in Fixed value
operation mode.
Change to Fixed value operation
mode.
Controller does not equilibrate to
program set-points.
Controller is not in program oper-
ation mode, or program delay
Start the program again. If appro-
priate, wait for the program delay
Program duration longer than
programmed.
Tolerances have been pro-
grammed.
For rapid transition phases, do
NOT program tolerance limits in
order to permit maximum heating,
refrigerating, or humidification