Issue 03/2019 Art. No. 7001-0026
Operating Manual
Translation of the original operating manual
BD / BD-UL (E2) Incubators with natural convection
ED / ED-UL (E2) Drying and heating ovens with natural convection
FD / FD-UL (E2) Drying and heating ovens with forced convection
with microprocessor temperature controller
Modell Model version Art. No. Modell Model version Art. No.
BD 23
BD023-230V 9010-0187
ED 240
ED240-230V 9010-0101
BD023UL-120V 9010-0189 ED240UL-208V 9010-0167
BD 53
BD053-230V 9010-0081
ED400
ED400-230V 9010-0075
BD053UL-120V 9010-0179 ED400UL-208V 9010-0168
BD 115
BD115-230V 9010-0088
ED720
ED720-230V 9010-0076
BD115UL-120V 9010-0181 ED720UL-208V 9010-0169
BD 240
BD240-230V 9010-0095
FD 23
FD023-230V 9010-0194
BD240UL-120V 9010-0183 FD023UL-120V 9010-0196
BD400
BD400-230V 9010-0073
FD 53
FD053-230V 9010-0082
BD400UL-120V 9010-0176 FD053UL-120V 9010-0128
BD720
BD720-230V 9010-0074
FD 115
FD115-230V 9010-0102
BD720UL-120V 9010-0177 FD115UL-120V 9010-0129
ED 23
ED023-230V 9010-0190/0191
FD 240
FD240-230V 9010-0104
ED023UL-120V 9010-0192/0193 FD240UL-208V 9010-0130
ED 53
ED053-230V 9010-0078/0079
ED053UL-120V 9010-0131/0132
ED 115
ED115-230V 9010-0096/0097
ED115UL-120V 9010-0164/0165
BINDER GmbH
Address: Post office box 102, 78502 Tuttlingen, Germany Phone: +49 7462 2005 0
Fax: +49 7462 2005 100 Internet: http://www.binder-world.com
E-mail: info@binder-world.com Service Hotline: +49 7462 2005 555
Service Fax: +49 7462 2005 93 555 Service E-Mail: service@binder-world.com
Service Hotline USA: +1 866 885 9794 or +1 631 224 4340 x3
Service Hotline Asia Pacific: +852 390 705 04 or +852 390 705 03
Service Hotline Russia and CIS: +7 495 988 15 16
DOMINIQUE DUTSCHER SAS

Table of Contents

Questions and Answers

  • J
    Jason BrownAug 17, 2025
    Why is my Binder Climate chamber not working?
    • C
      Christine MayerAug 17, 2025
      Your Binder Climate chamber might not be working due to several reasons. First, check the connection to the power supply. Second, the chamber fuse might have responded, so check it and replace it if needed. If it responds again, contact BINDER service. Finally, the controller could be defective, in which case you should also contact BINDER service.
  • A
    Andrew Jimenez Jr.Aug 19, 2025
    What to do if Binder 9010-0082 set-point temperature is not reached?
    • N
      Nancy YoungAug 20, 2025
      If the set-point temperature isn't reached in your Binder Climate chamber, ensure the chamber door is completely closed. Also, inspect the door gasket and replace it if it’s defective. Calibrate and adjust the controller. Finally, check the power supply to ensure the voltage is correct (115V or 230V).
  • J
    Joanna ThompsonAug 22, 2025
    Why Binder Climate chamber doesn’t heat up but controller display working?
    • F
      fosterannetteAug 22, 2025
      If your Binder Climate chamber isn’t heating up, and the red heating control light is off but the controller display is working, it could be due to the timer being run off. Program the timer or switch to continuous operation. Another cause might be a defective semiconductor relay, in which case you should contact BINDER service.
  • S
    Sarah AguilarAug 25, 2025
    What to do if Binder 9010-0082 isn't working and the red alarm light is on?
    • J
      Janet HamiltonAug 26, 2025
      If your Binder Climate chamber isn't working and the red alarm pilot lamp is lit, it could be that the safety device class 2 has turned off the chamber. Let the chamber cool down and then press the RESET button. Also, the safety device class 2 might be defective, in that case contact BINDER service.
  • T
    Tony MirandaAug 27, 2025
    Why is my Binder 9010-0082 heating permanently?
    • A
      Ashley BarrettAug 28, 2025
      If your Binder Climate chamber is heating permanently and not maintaining the set-point, the controller may be defective, in that case contact BINDER service. Alternatively, the controller might need calibration and adjustment.
  • S
    Sheri ZimmermanAug 31, 2025
    How to fix high temperature and red alarm on Binder Climate chamber?
    • S
      stephaniebrownAug 31, 2025
      If the temperature inside your Binder Climate chamber is too high and the red alarm light is on, the safety device class 3.1 may have been triggered. Check the settings of the temperature set-point and of the safety device class 3.1.
  • Y
    yorktinaSep 2, 2025
    Why does my Binder Climate chamber take longer to heat up?
    • S
      Sara BurgessSep 2, 2025
      If you're experiencing deviations from the indicated heating-up times with your Binder Climate chamber, it might be because the chamber is fully loaded. Try loading the chamber with fewer items or allow for longer heating-up times.
  • K
    Kaitlyn EwingSep 4, 2025
    Why is my Binder 9010-0082 in stand-by mode?
    • E
      Eric CervantesSep 5, 2025
      If your Binder Climate chamber is without function and only the green 'stand-by' LED is lit, the chamber is likely in stand-by mode. Press the ON/OFF button until the display lights up.
  • L
    ldavisSep 8, 2025
    What to do if Binder 9010-0082 fan is not working?
    • H
      Heather MclaughlinSep 8, 2025
      If the fan in your Binder Climate chamber isn’t turning or is turning too slowly, the fan might be defective, in that case contact BINDER service.
  • M
    Michael SkinnerSep 12, 2025
    Why does my Binder 9010-0082 controller return to normal display?
    • T
      Tracy LawrenceSep 12, 2025
      If the controller on your Binder Climate chamber returns to the Normal Display from any level, it's likely because no button was pressed for approximately 30 seconds. Repeat your entries, entering the values more rapidly.

Binder 9010-0082 Specifications

General IconGeneral
BrandBinder
Model9010-0082
CategoryClimate chamber
LanguageEnglish