88
IEC 61000-4-8
Note: U
T
is the a.c. mains voltage prior to application of the test level.
Electromagnetic Immunity
The Crossvent is intended for use in the electromagnetic environment specified below. The customer or the user of the
Crossvent should assure that it is used in such an environment.
Immunity Test Compliance Level Electromagnetic Environment Guidance
Conducted RF IEC
61000-4-6
Radiated RF IEC 61000-
4-3
6 V rms i
10 V/m
The Crossvent is suitable for the electromagnetic
environment of typical commercial or hospital settings.
The Crossvent was also tested for radiated immunity to RF wireless communication equipment at the test levels below.
Frequency (Hz) Modulation Level V/m
385
Pulse, 18 Hz, 50% DC
27
450
FM, 1 kHz Sine, ±5 Hz Deviation
28
710, 745, 780
Pulse, 217 Hz, 50% DC
9
810, 870, 930
Pulse, 18 Hz, 50% DC
28
1720, 1845, 1970
Pulse, 217 Hz, 50% DC
28
2450
28
5240, 5500, 5785
9