biodex 950-192 - Spasm Control; Emergency Stop

biodex 950-192
48 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
21 Biodex Medical Systems, Inc. © 2017
Spasm Control
For user safety, the medBike is equipped with a Spasm Control sensor. Medbike spasm control
detects sudden twitching and/or rapid pedal movement in opposite directions. If this occurs,
pedal movement will gently cease and the screen will display an “Overload!” message. A
therapist is able to intervene and decide what action to take.
Figure 2.16. Spasm control warning screen.
NOTE: If pedal rotation is completely resisted in the Passive Mode for a period of time (about a
minute), you may see a “Motor over current protection” message. To protect the motor
mechanism, the device will suspend operations. When this event occurs, you must turn the
device off with the ON / OFF switch at the base, and then turn the power back on.
Emergency Stop
The medBike is equipped with an EMERGENCY STOP Button. If, for any reason, a user wants to
quickly pause or discontinue an exercise session be it in Active or Passive Mode it can be
accomplished by pressing the STOP button. To continue exercising, the STOP button must be
pressed a second time. You will have the option to resume the session you were in before
pressing the STOP button or finishing the session and starting a new one.