8. Maintenance and Service
The BioStep 2 has been designed to provide many years of dependable use. To help
ensure that this product performs according to maximum specifications and to increase
the life of the product, please note the following general cleaning instructions and
maintenance procedures:
Daily Maintenance
NOTE: Do not use cleaning solutions containing ammonia.
1. As required, clean all exterior surfaces, upholstery and restraining straps.
Specialized vinyl
cleaners or protectants are recommended for upholstery and
cushions. Otherwise, use a solution of warm water and mild detergent.
2. As needed, wipe the display using a soft rag dampened with alcohol.
Annual Maintenance
1. Check to be sure all functions operate properly.
2. Inspect the seat for stability.
3. Check for any signs of wear.
4. Check for noisy components.
Diagnostic Mode
Certain information and various tests are available only from Diagnostic Mode. To view this
information and run the diagnostic tests, follow the steps below.
1. Begin pedaling to power-up the BioStep 2. Once the display panel lights, you do not
have to continue pedaling.
2. To enter Diagnostic Mode, press the Scroll/Pause Button and the Weight Button at the
same time.
3. Press the Scroll/Pause Button by itself to scroll through the following information:
Upper Window Lower Window
firmware release version (release number)
total number of hours of operation HRS
LOAd - used to test the resistance mechanism 0 - increase for test purposes
EE - electrically erasable PROM 055
Atod - value of A to D converter indicates how
much current is being passed through BioStep 2 0
LED 8.8.8.8 - flashes from upper to
lower window and to upper
window again to test for
defective segments
4. To exit Diagnostic Mode, either:
a. Press the Scroll/Pause Button again, or
b. Leave the BioStep 2 in Diagnostic Mode for one minute (whether or not anyone is
exercising).
You will hear a long beep and the display will default to the power-up condition.
MAINTENANCE AND SERVICE — 20 —