10. Electromagnetic Compatibility
Electromagnetic compatibility tables for both BioStep 2 models are provided in this section.
BioStep 2
Standard Test Method Range Limits Result
IEC 61000-3-2 Harmonics
100 Hz to 2KHz
Class A
THD = 131.54%
IEC 61000-3-3 Flicker observation time (TP) 10 min
- -
max voltage change (dmax)
4% 0.00 %
max Rel steady state voltage
3.3% 0.00 %
change (dc) duration of d(t)>3%(t)
.2 sec
0.00 sec
short term flicker Sev (PST)
1.00 0.00
long term flicker Sev (PLT)
.65
0.00
IEC 61000-4-2
Electrostatic
Contact: 2, 4, & 6Kv pos/neg
no degradation complied
Discharge
1pps for 10 sec
of performance
Air: 2, 4, & 8Kv pos/neg
1pps for 10 sec
IEC 61000-4-3 Radiated Immunity
80 MHz to 1000 MHz / 3v/m
no degradation complied
Horiz & Vertical @ 2M
of performance
1000 MHz to 2500 MHz / 3v/m
Horiz &Vertical @ 1M
IEC 61000-4-4
Electrical Fast
PWR Input leads .5, 1, & 2 no degradation complied
Transient/Burst, Kv / pos & neg / 5KHz Rep Rate
of performance
Power Leads complied
IEC 61000-4-6
Conducted Immunity,
Freq Range 150 KHz to 80 MHz no degradation complied
Power Leads
/3Vrms Modulation 1khz, 80%, AM
of performance
IEC 61000-4-8 Magnetic Immunity
3A/M RMS @ 50/60 Hz
no degradation
complied
of performance
IEC 61000-4-11
Voltage Dips and Int
Pause
%
no degradation complied
Interrupts
Duration
Between Reduction of performance
Int
20msec
pause
>100%
no degradation complied
10 sec
of performance
Int
100msec
pause
60%
no degradation
complied
10 sec
of performance
Int
500msec
pause
30%
no degradation
complied
10 sec
of performance
Int 5000msec
pause
>100%
no degradation
complied
10 sec
of performance
CISPR 11 Conducted Emissions
150 KHz - 500 KHz / 500KHz - 5MHz
66-56 / 56/ 60
complied
Edition 4: 2003
/ 5 MHz - 30 MHz, Class B, Group 1
dbuV QP
56-46 / 46 / 50
dbuV AV
CISPR 11
Radiated Emissions
30 MHz - 230 MHz / 230MHz - 1GHz,
30/37 dbuVm @
complied
Edition 4: 2003 Class B, Group 1
10 m
IEC 61000-4-5
Surge Immunity,
.5 & 1 Kv, Differential 1ppm, pos/neg
no degradation complied
Power Leads
.5, 1, & 2Kv Common 1ppm, pos/neg
of performance
ELECTROMAGNETIC COMPATIBILITY — 22 —