– 21 –
Cleaning
The instrument must be cleaned in the following
situations to assure correct operation:
if the piston becomes diffi cult to move
after dispensing liquids which form deposits
before changing the reagent
prior to long term storage
should liquid have accumulated in the
closure cap
prior to sterilization
before replacing the valves
Warning!
Follow all Safety Instructions (see pages 4, 5). The
glass cylinder, valves, fi lling and discharge tubes
contain reagent. Point the valves and tube openings
away from your body. Wear protective clothing, eye
protection and appropriate hand protection.
Reinigung
Damit eine einwandfreie Funktion gewährleistet ist,
muss das Gerät in folgenden Fällen gereinigt werden:
wenn der Kolben schwergängig wird
nach Dosieren von Flüssigkeiten, die Ablagerungen
bilden
– vor Reagenzwechsel
– vor längerer Lagerung
wenn sich Flüssigkeit in der Verschlusskappe
angesammelt hat
– vor dem Autoklavieren
– vor Ventilaustausch
Warnung!
Alle Sicherheitsbestimmungen befolgen (s. S. 4, 5).
Glaszylinder, Ventile, Ansaugrohr und Dosierkanüle
sind mit Reagenz gefüllt. Öffnungen von Ansaug-
rohr, Dosierkanüle und Ventilen niemals auf den
Körper richten. Schutzkleidung, Augenschutz und
ausreichenden Handschutz tragen.
Hinweis:
Die Kolben der Geräte dürfen nicht vertauscht
werden!
Note:
Never change the pistons of the instruments!