1) Features
- LCD Graphic Display 800 X 480 dots, 3x1/3x2/6x1/6x2
- Channel with base line movement possibility
- Standard 6 channels
- Real time heart rate display
- Real time display and print out
- Various patient information
- (Patient Name, Sex, Age, Weight, Height)
- High Resolution Thermal Printer
- Defibrillator protection
- Rechargeable Battery Operation
2) Specifications
- Diagnostic Electrocardiography
󳢯
Input circuit : Floating input
󳢯
Leads : Standard 12 leads
󳢯
Input impedance : 20Mohm min
󳢯
Input circuit current : 0.05 A max
󳢯
Calibration voltage : 1mV with ± 2%
󳢯
CMR : 100dB or more
󳢯
Polarization voltage : ± 350mV min
󳢯
Time constancy : 3.2 seconds (0~+20%)
󳢯
Sensitivity : 5, 10, 15 mm/mV ± 5%
󳢯
Frequency response : 0.05 ~ 100Hz (13dB)
󳢯
Filters : AC50/60 Hz, -20dB or lower, Muscle tremor 35 to 45 Hz, -3db(-6db/oct)
󳢯
A/D conversion : 12bits
󳢯
Sampling rate : 2msec
3) Physical Specifications
- Power requirements : AC 100 ~ 240V, 50/60Hz
- Power consumption : 50VA(MAX)
- Battery: LI-Ion , 11.1V , 4400mAh
- Continuous operating time with battery : Max 4Hours
- Dimension : 327(W) X 206(L) X 87(H) mm (W x D x H)
- Weight : 1.75Kg(Incl. battery)
4) Interface
- RS-232C port
- LAN Port (Program upgrade)
- USB Port (Data storage)
5) Operating Environment
- Temperature : (Operating : 10 ~ 45 / Storage : -10 ~ 50)
- Humidity : (Operating : 30 ~ 85% / Storage : 20 ~ 95%)
- Atmospheric Pressure : (Operating : 70 to 106 kPa / Storage : 70 to 106 kPa)
6) Product specification listed on the manual are subject to change according to the our
design change procedure without notice.