70
DC Current: ± (% of reading + % of range), 23 °C ± 5 °C
Burden Voltage
(1)
& Shunt Resistor
5.0000 mA
50.000 mA
500.00 m A
5.0000 A
20.000 A
0.1 µA
1 µA
10 µA
100 µA
1 mA
5.1000
51.000
510.00
5.1000
21.000
(3)
0.05%+0.01%
(2)
0.05%+0.008%
(2)
0.05%+0.008%
0.25
%+0.01%
0.25
%+0.01%
<0.6 V / 100 Ω
<0.06 V / 1 Ω
<0.6 V / 1 Ω
<0.1 V / 10 mΩ
<0.6 V / 10 mΩ
(1)
Typical voltage across the input terminals at full scale reading.
(2)
Use REL function
(3)
In 20 A range,>10-20ADC is readable for 20 seconds maximum
Maximum Input and Overload Current Protection: 1A/ 250V fuse.
AC Current (True RMS, AC Coupling)
ACCURACY: ± (% of reading + % of range), 23 °C ± 5 °C
Accuracy(1 year)
(1)
23 °C ± 5 °C
5.0000 mA
50.000 mA
500.00 m A
5.0000 A
20.000 A
1.5%+0.16%
1.5
%+0.16%
1.5
%+0.16%
2.0
%+0.16%
2.0
%+0.16%
0.5%+0.08%
0.5
%+0.08%
0.5
%+0.08%
0.5
%+0.1%
0.5
%+0.1%
2.5%+0.16%
2
%+0.12%
2
%+0.12%
------
------
(1)
Specifications are for sine wave inputs >5% of range.
In 20 A range,>10-20Aac is readable for 20 seconds maximum
Measurement method: True RMS measurement
Maximum Crest factor: 3.0 at full scale
Maximum Input and Overload Current Protection: 1 A/ 250 V fuse.
Burden Voltage
(1)
/shunt Resistor
5.0000 mA
50.000 mA
500.00 m A
5.0000 A
20.000 A
0.1 µA
10 µA
100 µA
1 mA
10 mA
5.1000
51.000
510.00
5.1000
21.000
(2)
<0.6 V / 100 Ω
<0.06 V / 1 Ω
<0.6 V / 1 Ω
<0.1 V / 10 mΩ
<0.6 V / 10 mΩ
(1)
Typical voltage across the input terminals at full scale reading.
(2)
In 20A range,>10-20Aac is readable for 20 seconds maximum
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com