Operation Modes 35
6.1.3 Constant Current Mode (CC)
In constant current mode, the load will sink a current in accordance with the programmed value regardless
of the input voltage.
Figure 6.1 Constant Current Mode
To run the CC operation RUNMODE must be set to Normal and CC must be selected by pressing the
button. For more information about RUNMODE please refer to section RUNMODE.
Congure CC Parameters
The following parameters are congurable under the Setup menu once CC mode is selected.
Range
Denes the maximum allowed current set value. Use this limit to protect against accidentally entering
excessive current values either from the keypad or dial. Please refer to section Range Conguration.
UP/ DOWN
Sets the rise and fall slew rate of the load, which determines the rate at which the input current increases
and decreases to a new programmed value. Please refer to section Slew Rate Conguration.