EasyManuals Logo

BK Precision 8600 Series User Manual

BK Precision 8600 Series
81 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #42 background imageLoading...
Page #42 background image
42
Parameters
VOLTage | CURRent | TWO
*RST Value
TWO
Examples
TRAC:FEED VOLT
Query Syntax
TRACe:FEED?
Returned Parameters
<CRD>
Related Commands
TRAC:POIN
TRACe:FEED:CONTrol
This command is used to select the buffer control. With NEVer selected, storage into the buffer is
disabled. When NEXT is selected, the storage process starts, fills the buffer and then stops. The buffer
size is specified by the :POINts command.
Command Syntax
TRACe:FEED:CONTrol <CRD>
Parameters
NEVer | NEXT
*RST Value
NEVer
Examples
TRAC:FEED:CONT NEXT
Query Syntax
TRACe: FEED:CONT?
Returned Parameters
<CRD>
Related Commands
TRAC:FEED
TRACe:DATA?
When this command is sent and the electronic load is addressed to talk, all the readings stored in the
buffer are sent to the computer.
Query Syntax
TRACe:DATA?
Returned Parameters
{<NR3>, <NR3>…etc,} or {<NR3> <NR3>, <NR3> <NR3>
…etc} (if TRACe:FEED TWO is set)
TRACe:FILTer
This command is used to select whether the data in buffer is filtered data.
Command Syntax
TRACe:FILTer[:STATe] <BOOL>
Parameters
0 | 1 | ON | OFF
*RST Value
OFF