:ERRor? Query (read) Error Queue.
:CLEar Clears messages from the Error Queue.
:LOCal
Take load out of remote and restore operation
of front panel controls.
:REMote Place load in remote control.
:RWLock Lockout front panel controls.
STATus Command Summary
Commands Description Default
:STATus (note 1)
:OPERation Path to control operation status register.
[:EVENt]? Read the event register. (note 2)
:ENABle <n> Program the enable register. (note 3)
:ENABle? Read the enable register.
:CONDition? Read the condition register.
:QUEStionable Path to control questionable status register.
[:EVENt]? Read the event register. (note 2)
:ENABle <n> Program the enable register. (note 3)
:ENABle? Read the enable register.
:CONDition? Read the condition register.
:PRESet Return status register to default states.
Notes:
1. Commands in the STATus subsystem are not affected by *RST and :SYSTem:PRESet. The effects of
cycling power, *CLS and :STATus:PRESet are explained by the following notes.
2. Event Registers: Power-up and *CLS clears all bits of the registers. :STATus:PRESet has no effect.
3. Enable Registers: Power-up and :STATus:PRESet clears all bits of the registers. *CLS has no effect.