EasyManua.ls Logo

Blue connect - Probe Replacement; Probe Lifespan and Care; Probe Replacement Steps; Step 3: Remove Cap and Nut

Blue connect
21 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
11
Blue Connect
Blue Fit50 » USER MANUAL
C. Probe replacement
The probe needs to be replaced approximately every two years
6
. Please contact your reseller to get a
new one.
NB: To ensure maximum longevity of the probe, keep it for as long as possible in a humid
environment. If you take the device out of the water, you must screw it to the storage cap with KCl
(saturated potassium chloride), pH4 or, as a last resort, pool water.
The box includes:
Installation:
In “Settings“ » ”Blue Connect, press “Probe replacement“.
Take out Blue Connect from the water. Wipe the device and start the procedure in a dry,
clean place.
Remove the usage cap and nut. If you struggle to unscrew the nut, use adjustable pliers by
placing them on this exact spot (Image A).
Image A:
6
Make sure not to leave the probe out of the water without the storage cap lled with KCl (saturated potassium chloride), pH4
or, as a last resort, pool water.
1.
2.
3.

Other manuals for Blue connect