Mountingandcommissioning
©byBlumNovotest
15
V2A|04/2018
1. Cleanthemea suringsystemcarefully.
2. Untightenthescrewsatthebatterycompartmentlid.
3. Pullbackbothscrewsandremovethebatterycompartmentlid.
4. Removethebatteries.
5. Cleananddrythebatterycompartmentcarefullywithabsorbentpaper.
6. Putinnewbatteries.Pleaseconsiderthepolarityofthebatteries.
7. Pleasecontrolifthegasketisinsertedpr
ofessionally:
smal
l,evenlyspacebetweensealinglipandwall
8. Putinthebatterycompartmentlidpreciselyandmountit(M
d
=0.7Nm).Thereby,please
checktheOringsandgreaseit,ifnecessary.
9. Basicadjustmentsaredisplayedautomatically,seechapter4.6.