V2A|04/2018
3.5 Technicaldata
Tab.32
TC50 TC51 TC5330/
TC76T
Approachdirection ±X,±Y, ‐Z XY
a
,‐Z,+Z
a
viaspindleindexing
±X,±Y, ±Z
b
b
seechapter4.5
Measuringforce
•directionX/Y: seeTab.35...Tab.37
Measuringforce
•directionZ:
7.0 N 4.0 N 5.9 N
2.0 N
Max.deflection
•directionX/Y(seeTab.34): ± 15° 15°
a
±15°
•directionZ: 10 mm Z:6mm
+Z:3mm
5mm
Max.acceleration 50 m/s
2
100 m/s
2
50 m/s
2c
c
withaccessories(e.g.extension,angle):max.acceleration10m/s
2
Repeatability
d
d
StylusL=50mm
0.3 µm 2σ 0.3 µm 2σ 0.4 µm 2σ
Max.probingspeed 3 m/min 5 m/min 2 m/min
Mass 925 g 920 g 710 g
(TC5330)
Protectionclass IP68
Signaltransmission Infrared
Transmitterrange 50° ... +70°inZ,360°inX/Y
Battery(2pieces) SaftLithiumLS14500 (AA, 3.6 V) 2600 mAh
ToolholderHSKSKBT BTH50
Storagetemperature 20 °C ... +70 °C
Operatingtemperature +5 °C ... +50 °C