Mountingandcommissioning
©byBlumNovotest
17
V2A|04/2018
1. Insertthemeasuringsystemintothetoolholder(systemBTH).
2. Tightentheclampingscrewjustslightly.
3. Inserttheprobeintothespindle(TC50,TC52,TC5310,TC5330,TC5410only).
4. Setthedialgaugeonthestylus.
5. Rotatetheprobeby360°.
6. Adjusttheconcentricity(4hexagonsocketscrew s2mm):
e.g.Xdir
ection:
Releasethecentringscrews(2a)and(2c)justslightly.
Adjusttheconcentricitybyalternatereleaseandtighteningofthecentringscrews(2a)and
(2c).
Ydirection:
Releasethecentringscrews(2b)and(2d)justslightly.
Adjusttheconcentricitybyalternatereleaseandtighteningofthecentringscrews(2a)and
(2c).
Concentricitydeviation:
TC50,TC52,TC
5310,TC
5330,TC5410:<10µm
TC51:<100µm
Allcentringscrewsaretightened.
7. Tightentheclampingscrew.
Theconcentricityadjustmentisnowfixed.
8. Checktheconcentricity.
9. Carryoutacalibrationcycle.
AdjustmentoftherunoutatTC51withoptic alpre set ter!